Test Kit is for In-vitro diagnostic (IVD) use to determine C-reactive protein (CRP) quantitatively from human blood and serum/plasma. It can be user to diagnosis and follow-up for patients with infectious and non-infectious inflammatory diseases. 


  • Wide measurement range  
  • 'Vertical-flow' immunoassay 
  •  Point-Of-Care testing

Specifications

Simple Procedure

  Principle
Flow-through immunometric assay
  Sample Type
 Serum/plasma, Whole blood (capillary and venous)
  Sample Volume
 5 𝛍L
  Test Result
 CRP (C-reactive protein)
  Test Range
 2.5 - 200 mg/L
  Precision
CV < 5%
  Test Time
Less than 4 minutes
  Calibration 

Olympus AU 5800, 

ELITECH Selectra ProM

Performance

Precision
Within-run
Day-to-day
Lv.1
Lv.2
Lv.3
Lv.1
Lv.2
Lv.3
CRP (mg/L)
2.83
46.41
111.08
2.95
45.31
108.99
SD (mg/L)
0.07
0.82
2.53
0.10
0.76
3.35
CV (%)
2.45
1.77
2.27
3.38
1.67
3.07

Contents

Name
Quantity
Package
Remark
TC/Test Cartridge
50 x 1
Individual pouch
Disposable
R1/Reagent
50 x 1
Zipper bag
Disposable
R2/Conjugates
1
Bottle
For 50 tests
R3/Wahing Solution
1
Bottle
For 50 tests
IC/Item Card
1
-
-
IFU/Instructions for use
1--

Test Principle